Physics of Semiconductor Devices

From Advanced Projects Lab
Revision as of 15:55, 14 June 2015 by Wikiuser (talk | contribs)
Jump to: navigation, search

Physics of Semiconductors Project

Physics of Semiconductors Web Page

The goal of this project is to characterize different semiconductors using various techniques. Currently, PIN diodes and Schottky diodes have been tested using various methods. The existing methods can be improved, new methods can be implemented, and new diodes could be tested to further expand this project.

Group 1

One goal of the Physics of Semiconductor Devices project is to measure captured carriers within the band-gap of a semiconductor. If a carrier (a hole or electron) is captured and then released in the band, the capture center is called a trap. These traps arise from defects or impurities in the material. One technique for measuring these traps is through Deep Level Transient Spectroscopy, or DLTS. This technique essentially measures the change in capacitance of the semiconductor.

We are using a high‐frequency capacitance transient thermal scanning method to observe traps in the depletion region of our diode. The capacitance transients can give us information about trap concentration, activation energy and electron‐ and hole‐capture cross sections for each trap. These properties are of interest to industry.

Our circuit is built on a copper plating encased in a steel container. This is to reduce stray capacitance and noise from the high frequency signal (our signal is 30 MHz.) The plating provides a large common ground that will reduce these effects. In the future copper tape soldered near the input and output ports will further reduce these unwanted effects. Our diode is a schottky diode, which is used for it's fast switching properties.

The premise behind DLTS is to use a short pulse to momentarily forward bias a diode that is in a steady state of reverse bias. This short pulse will inject carriers (in our case electrons) into the depletion zone of our semiconductor (putting the diode in forward bias will shrink the depletion zone and allow electrons to enter). This will change the capacitance of the diode. Semiconductors are characterized by their unique conductive behavior. A semiconductor has a fermi energy that lies inside a band gap in electron energy levels. This band gap is small in a semiconductor however, allowing electrons to cross the band gap at relatively low temperatures (as compared to an insulator.) The rate at witch electrons are emitted from a trap is what will characterize the capacitance transient. The emission rate is proportional to a boltzman factor and depends exponentially on the energy difference between the trap level and the conduction band. In essence we will inject the depletion zone with electrons, thus raising the capacitance of the diode. Some electrons will become trapped. Thermal energy will eventually kick these electrons out of the traps and back into the charged portion of the semiconductor. The capacitance will spike when the electrons are introduced and slowly return to normal as the electrons are kicked out the traps. If the traps were not present we would expect a linear return to steady state capacitance.

Our circuit will mix a known voltage signal with another voltage signal passing through the diode. We have created a circuit that will split the incoming signal. On one leg of our circuit the voltage signal will run through the diode. The other leg consists of a 180 degree phase shifter that will recombine with the signal coming out of the diode, thus canceling the signal and outputting a signal of 0 amplitude. In reality we cannot exactly cancel out the signal coming out of the diode, so after the phase shifter we have a variable capacitor. This capacitor allows us to very minutely shift the phase of the signal. This will allow us to fine tune the phase of the signal we are using to cancel the signal coming out of the diode. At the end of the circuit both legs are attached to a variable resistor. There are voltage drops across both legs of the circuit and the variable resistor allows us to change the amplitude of the signals to find the ideal for cancellation. By adjusting the variable resistor and capacitor an input signal measured at 248 mV resulted in a 1.6 mV output.

The original signal will input into a mixer with the signal coming out of the circuit. The signals are multiplied in the mixer and the resulting signal will be output on an oscilloscope. During steady state the output coming out of he circuit should be close to 0 so no signal will be measured. When we introduce the pulse the signal on the bottom leg of our circuit will no longer cancel the signal coming through the top wave and a signal of interest will be measured. This will result in a spike of voltage that will gradually return to zero as the electrons are kicked out of the traps in the depletion zone of the diode. We will observe a voltage transient and from that we will be able to calculate the capacitance transient we are looking for.

As stated earlier, thermal energy is what releases electrons from the traps. Electrons are released faster as the temperature increases, thus the capacitance transient is "thinner" with respect to time for higher temperatures (it takes less time for the capacitance to return to zero). To make measurements we will

consider emission rates which are what define the capacitance transient. We can define emission rate "windows", a certain rate in the capacitance transient, to measure only emission rates that we define. For each window we can take measurements at many different temperatures and we will see peaks at temperatures (because emission rate is temperature dependent) that match the window we defined. If we change the window and repeat the measurement we will see peaks at different temperatures. If we plot the width of the peaks against 1/Temperature we will observe a straight line, the length of which will be the rate window we defined. By measuring several different windows at varying temperatures we will observe a long straight line. The slope of that line is the activation energy for the trap. We will use a cryostat with liquid nitrogen to measure the diode at different temperatures.


Group 2: Zach Small and Chris Luetjen

Figure 1

For a standard diode its characteristics depends on what material the diode is made out of and with what it has been doped with. For the 1N4004, we measured an I-V characteristic as shown in Figure 1.

This is exactly what we would expect, as Diode Incorporated; the manufacturer, gives us the characteristic graphs of these diodes 1N4004_spec_sheet. Note that in the graph, the first point after point (0, 0) is the next none zero point, meaning at the current remained at 0 until 5.2 volts across the diode.

The standard diode will have a small amount of capacitance current flows across it. This capacitance is called the Junction Capacitance and is because charge is stored inside the diode and a small amount will build up. Figure 2 is C-V characteristic of the P6KE6.8A diode, of which this is the spec sheet: P6KE_spec_sheet. This data was taken using a voltage divider, shown in Figure 3

Figure 3

and the voltage divider equation, . f=10 Hz, Vin=20 mV (AC Voltage peak to peak), R=1 kohm, and bias Voltage=-500 mV to 1200 mV was chosen to take this data. As you can see, when reverse biased the voltage was basically zero, this is because the electrons and holes drift in the opposite direction and leave their respective n or p type region.

Figure 2

Carrier Dynamics

Carrier dynamics are the dynamics between the charge carriers in the semiconductor and how they interact with each other and external factors.

Depletion Region Width

The setup shown in Figure 4 was used to measure the carrier dynamics of a laser diode. For this experiment, the GH0781JA2C diode was used, the spec sheet can be found here, GH0781JA2C_spec_sheet. We used the two lenses to expand and collimate the lens. This is important, because the microscope objective, which is focusing our beam down to a point in order to shine on the semiconductor of laser diode, is designed to be fully filled of light in order to work properly. The chopper was used to chop the light into a frequency in which other parts of the room where not emitting radiation at. This frequency was then sent to the lock-in amplifier, which amplifies the signal at the reference frequency. This helps separate our data, the voltage over the laser diode, from the noise in the rest of the room. Before using the lock-in amplifier, we experienced trouble getting readings on the multimeter, which was measuring unchopped light. To fix this, we connected our diode with a coaxial cable instead of just plain wires. This was to protect the wires, which were acting as antennas and were picking up frequencies from other electrons in the room, so the lights, refrigerators, etc. This did fix the problem, but we still received significant noise, and thus used the lock-in amplifier. We used the 7280 whose instruction Manuel can be found here: 7280_spec_sheet_and_instruction_manuel The laser diode was also reverse biased. This is because there would be no voltage or current across the laser diode, and therefore the only voltage we would read would be that from focusing our IR diode laser (The New Focus 6224 was used) at 784 nm on the laser diode. The purpose was to measure the depletion zone of the diode, which we measured at -1 V, 0 V, and 1 V. The data is shown in Figures 5, 6, and 7 and is what we expected. The depletion region increases the more negative the bias across the diode. This is because the carries, electrons and holes, are evacuating the junction at negative bias, and thus the region with no carriers becomes larger.

Although our plots display the correct trend between bias voltage and depletion region width, the scale is off. We would expect the depletion region to be roughly 1 micron, and ours measures 30-50 times that. This discrepancy is most likely due to the beam size. Our data assumes a beam size of 1 micron, which is unachievable with our system. The best beam size possible for our set up would be 1.5 microns, assuming the beam incident on our microscope objective is perfectly Gaussian. Realistically our beam size is roughly 10 microns, but a measurement should be made verify this. This measurement could be made by using the knife-edge technique, or a high resolution beam profiler or CCD camera.

Figure 4
Figure 5
Figure 6
Figure 7

Band Edge

From quantum mechanics we know that atoms exist in discrete energy levels. When N atoms are brought together, such as in a crystal, the atomic energy levels interact. This interaction causes the energy levels to split into N distinct levels. However, in crystal that contains a large number of atoms, there are so many levels, that they are indistinguishable from one another, and what is known as an energy band is created. In this band, electrons can move freely since the neighboring bands are almost touching.

When working with semiconductors, the atoms are arranged in crystalline structures, so the aforementioned energy bands are present. However, practically only two of the bands are relevant, the conduction band and the valence band. Specifically, the distance between these bands, called the band gap. In order for a semiconductor to conduct electricity, there must be electrons in the conduction band, but no electrons can exist in between the valence and conduction band, so the electrons must be excited by one “burst” of energy, such as from a photon or heat. There is also a momentum component to this transition, however, it will be omitted since we are assuming it holds no bearing in our semiconductor, which is intended to have a direct band gap.

Since a minimum energy is required to drive an electron from the valence band to the conduction band, when a photon is absorbed by a semiconductor, it can only cause this transition if it has an energy greater than or equal to the band gap. And since a photon energy is related to the wavelength by the equation, E=hc/λ, certain wavelengths can cause an electron to by driven up into the conduction band. By varying the wavelength of light incident on a semiconductor, it is possible to determine the band gap. To measure how many electrons are excited into the conduction band, a small bias voltage is applied to the semiconductor. This causes any electron in the conduction band to flow out towards the bias, creating a current which we can then measure.

With the same setup that was used for measuring the depletion region we took band edge measurements. This was done by varying the wavelength of the tunable IR laser used in the carrier dynamics setup and directing the beam straight at the center of laser diode. The data found was conflicting and as seen in Figure 8, it varied a bit and flat lined at certain wavelengths. To fix this we added in a isolator and a wave plate that adjusted the polarity of our beam to that which the isolotor would allow to pass through. This was to remove any back reflection that may have been skewing our data. This did not fix the problem, so we took a spectrum of our laser at each wavelength, varying from 784 to 804 nm. As seen in Figure 9, the laser we were using does not tune past Figure 6 and at one point emits two wavelengths when it should be emitting one. For future band edge measurements, a different laser will have to be used.

Figure 8
Figure 9


Deep Level Transient Spectroscopy

The Deep Level Transient Spectroscopy was discussed in detail above by Group 1


Electronics

Figure 10
Figure11
Figure12

In Figure 10 is a picture of the circuit that they constructed in Figures 11 and 12 is the schematics for the circuit. We utilized this circuit to make measurements on three different diodes, a standard diode; P6KE , a Schottky Barrier Rectifier; 1N5817_1.0A , and a Fast Response diode; 1N4148. The links are to the spec sheets of said diodes. The measurements are at 1 V and 2.3 V. You will notice that the decay curve changes from a negative decay to a positive decay and this is when the sum of the offset and the pulse is greater than the threshold voltage of the diode.

Current Data


The data above shows a decay rate for what we originally thought was the decay of carriers through the diode. It turns out that with no diode, a very similar decay is observed, although more slowly. It is possible to get almost the identical curve by varying the resister in the DLTS circuit. In Figure 13 a pulse is showed with no diode. It is shallower, but for it be to what we would expect it to be, it should rise and fall sharply with the pulse. It should look almost identical to the pulse sent in, however we do get a slight decay rate. Another indicator that the above data may be measuring something other than the decay rate of carriers is the decay across the pulse. We would expect it to remain flat with the pulse, as steady current passes through the diode at that point, meaning that carriers are flowing through the diode. To check if our circuit was working, we rebuilt the total circuit shown in Figure 11 with stand alone components and attached a standard diode. We then placed this same diode into our DLTS and measured the results of each, in which we observed the same result. The stand alone setup result is shown in Figure 14 and the DLTS is shown in Figure 15. In conclusion this data may be measuring the decay rate of carriers in the diode as it does change with the diodes, but we have inconclusive evidence until the pulse passing through the system is analysed.

Figure 13
Figure 14
Figure 15

Cryostat

Figure 16

As previously mentioned by group 1, a cryostat can be used to cool the diodes down to liquid nitrogen temperatures (77 K). By cooling the diodes, the thermal energy decreases significantly, so the ability for the charge carriers to escape the traps decreases, resulting in a slower voltage decay after the pulse.

To cool the diode we used a cryostat. A cryostat is essentially a liquid nitrogen Dewar with a chamber enclosed that has a plate in thermal contact with the liquid nitrogen. A picture of the cryostat is shown in figure 16. Everything in the chamber is cooled significantly, and anything in thermal contact with the plate, is also in thermal contact with the liquid nitrogen. We mounted the diodes in the cryostat by soldering them to a circuit board, and then affixing the circuit board to the thermal plate in the cryostat with the diodes all in contact with it. A schematic of this circuit board can be seen in figure 17. The diodes were then wired into a port so they could be connected to the rest of the circuit. The PIN layout of the port is shown in Figure 18.

Figure 17
Figure 18

The cryostat was never actually tested due to issues with the DLTS setup. And a new port should be made to connect the diodes in the cryostat to the external circuit, since the current port has brass pins on the outside, which solder does not stick to very well. That said, the current port will work fine if a hook or clip is attached to the each individual pin. The easiest method would be to use a BNC port.


Here is the Power Point Presentation of our final presentation. It includes a few pictures of our setup and the diodes and may prove useful if there are any questions in the actual physical setup of the experiment. Final Presentation.pdf







Group 3: Derek Hallman, John McGonigle, Alex Schachtner

We performed many of the above experiments as well, beginning with measuring the IV Curve of a diode, We used a PK6E30A diode, and used a current limiting circuit in order to limit the current through the diode. We varied the voltage in to measure the resulting current out and plotted the two. We performed this measurement for forward and reverse bias to begin understanding the pn junction and depletion region. We understand the mechanisms of charge carriers and how a diode works in forward and reverse bias.

IV Curve Forward Biased Diode
IV Curve Reverse Biased Diode

We then performed similar measurements while using a photodiode while injecting a photocurrent by shining a red laser pointer on the diode. We saw a similar IV curve for forward and reverse bias, however we never saw the reverse breakdown. While we saw a current effectively "turn on" similarly as the diode at around .6 volts in forward bias, though we never saw a reverse breakdown, even when taking the voltage up to 90-100 volts in reverse.

Photodiode Dark
Photodiode Light

Now we measured the capacitance of the diode as we understand that the depletion region will act effectively as a capacitator. Charges will migrate across the region until the resulting electric field will prevent further current. The smaller the depletion region, we found the higher the effective capacitance of the diode


Here our projects split where Alex pursued the DLTS experiments while John and Derek continued using laser diodes.

Measuring Depletion Region with a Laser

In this portion of the project we sought to utilize semiconductor properties to create light by stimulated emission of radiation by focusing a beam of near infrared signal onto a laser diode and then with piezo controls we scanned the beam across the active region in different directions and measured the depletion region.

We used an external cavity diode laser such that we could tune the wavelength of the laser and ensure that the diode was not mode hopping, such that the photons introduced into the laser diode were of constant energy. We wanted to select the most dominant mode for the greatest emission intensity. We used a wavemeter which measured the interference fringes of the HeNe beam and the beam from our laser diode. By steering a system of mirrors we were to align the beams at two points, ensuring that they were aligned as a line along the beam path. We used a bandpass filter with Near Infrared specs to allow the beam of our laser diode to pass and interfere with the HeNe beam, while disallowing the HeNe beam to reflect back into the laser diode which would alter the dynamics of emission and skew our beam.


Our set-up was as follows: Laser Diode -> Chopper -> Up/Down mirror -> Half-Wave plate -> Isolator -> Diverging lens -> Collimating Lens -> Microscope Objective -> Laser diode. [picture]

Our chopper singled out our beam's signal and in combination with the lock in amplifier we then began introducing our beam to the laser diode such that we could then scan the beam across the PN junction and the active region of the diode. The system of lenses were installed to adjust the spot diameter of the beam to the required length in order to fill the microscope objective properly which then projected the the signal onto the laser diode. Our goal here was to focus the beam down to .5um using Gaussian beam dynamics so we could scan the beam across the active region of the laser diode which was around 1um thick and 3-5 um across. We were also able to make more precise measurements when we used piezo controls in combination with the lock in amplifier to make our measurements.


Our data is as follows:

Travel X - Focal Axis Displacement
Travel Y - Horizontal Transverse Displacement
Travel Z - Vertical Transverse Displacement
Travel Y Lock-In
Travel Z Lock-In


Two Photon Upconversion - Derek Hallman

This leg of the experiment was Introducing a laser to the laser diode detector of wavelength one half of the wavelength that would normally cause single photon absorption. One photon promotes an electron to an imaginary state, second photon promotes energy to excited state, where each photon is roughly equal to half the energy of a photon that would excite an electron via one photon absorption. This is a Non-linear process (weak compared to OPA at low intensity, dominates at high intensity)

My experimental set-up is as follows:

I used an Erbium Doped Fiber Laser as a gain medium, pumped by a 980 nm laser to emit 1550 nm radiation. This radiation is roughly twice the wavelength of the photons that we used to stimulate single photon absorption, which made it suitable for two photon absorption. We used beam expansion similarly to single photon absorption, however we needed different lenses as the beam width from the free-space coupled EDFL was smaller than the External Cavity Diode Laser, and we wished to ensure the back aperture of the microscope objective was filled such that our calculations on beam waist would hold. We used a chopper for lock-in amplification to chop the beam so the Lock-In Amplifier could use the reference to eliminate noise, and we could really hone in on our signal.

Two Photon Set-Up
Erbium Doped Fiber Laser Pump and Decay
Erbium Doped Fiber Laser
Erbium Doped Fiber Laser Layout


My procedure is as follows:

I found that operating the pump laser at .500 amps that I could get an optical power of 17.92 mW, significantly greater than using an ECDL. Aligned two beams (1550, and ~640) using a flip mirror, and several irises so that I may roughly alight the focused beam from the microscope objective with the active region on the detector diode. Using the coarse controls on the translation stage, and finer adjustments using the piezoelectric displacement, I attempted to maximize photocurrent using single photon absorption, in order to roughly align the detector diode with microscope lens. It should be noted that the working distance with the visible light, and the IR light will be different, so further alignment with the IR beam is necessary. Thus I then reconfigured beam expansion for IR light, and worked to confirm that the 1550 beam was stimulating electrons via two photon absorption; I then worked to maximize the read-out on the lock-in amplifier. Confirming that two photon absorption is taking place, I begin scanning beam across diode surface.

Thus I began scanning beam across the surface of the laser diode detector, to understand two photon absorption as a function of displacement across the diode face I found that it was very difficult to find the active region of the diode using both two photon absorption and lock-in detection, and at the intensity I was working at only generated a signal in the order of magnitude of MicroVolts. To analyze the signal generated by scanning the beam across the diode, I took a 10x10 matrix, taking data in Microvolts at points that were a stepsize of about a micron (uM) apart. I then entered these matrices into MATLAB and wrote a function which both would interpolate these points and create various contour and surface plots of the matrices. This allows us a visual representation of voltage generated by Two-Photon Absorption of 1550nm IR radiation.

First Trial - Looks nearly Gaussian - Most likely due to poor beam focus
Second Trial - Showing a trend - Poor Data, likely scanning outside active region
Third Trial - Getting better

After collecting such poor data even while the lock-in amp seemed to be showing such clear gaussian trends upon displacement from what I believed was the active region, I resorted to reconfigure the translation stage, and used the coarse knobs to "hunt around" for the signal from introducing the beam to the active region. This was a very sensitive process as two-photon absorption requires high intensity. In doing so I found much higher voltages through the Lock-in amp than I had seen before - i.e. 10uV to about 30uV. I then collected the best data and what seems to be a clear indication of scanning the beam across the active region.

The best data is subject for further study, and possible publication. It is as follows:

Microvolts as a function of displacement
Microvolts as a function of displacement from a different angle
Surface plot of Microvolts as a function of displacement
Microvolts as a function of displacement - Best Results With Data Points

The data I have collected could have a variety of implications, and could provide the basis of future publications: - Temperature dependence - Increased penetration of IR photons into semiconductor sample - Different behavior across the PN junction from OPA

Two Photon Absorption - Derek Hallman & Evan Graham

We wished to improve upon measurements made and to make strides to better understand the nature of Laserdiodes and the physics of the semiconductors that comprise them. We saw in previous measurements how temperature could effect the behavior of charges in the laser diode. Our first goal was to introduce some temperature control to the design, so we replaced the diode that we have reverse biased for use as the detector. We introduced the LDM21 as a more advanced diode mount with integrated temperature control, we also replaced the diode in the mount from whichever we were using before (no exact specs, but we knew it had a bandgap ~785nm) with the thorlabs L785P6 diode which we then had the exact specifications for. We needed to correctly arrange the diode in the mount, and solder a D9 Connector so that we may monitor the signal out of the diode. We opted to use 2 BNC Coaxial cables as we used one of the tips of the cable as a ground, which grounded the diode, and the Cathode of the diode as we were operating in reverse bias / Cathode grounding, and the other BNC went to the anode so we could measure the signal. We monitored our signal using a LockIn-Amplifier and used Math Functions to monitor our signal as a potential difference (Anode - Ground / Cathode). Attempting to re-establish our signal we could not find evidence of two photon absorption using the 1550nm ~20mW laser we had used before while searching over the diode, and using OPA (~634nm ~2mw) to roughly align the pn junction with the microscope objective.

We then made some customizations to the Erbium Doped Fiber Laser that we had been using. We opted to use the pump of the EDFL to excite electrons as it both had a higher wavelength, and a much greater intensity. We created fiber connection terminals inside of the fiber laser so that we could still use the pump laser to excite the Erbium gain medium for using the fiber laser at 1550nm, or to an exterior port where we used the 980nm laser directly out. We noticed that when using the 980 wavelength, at the same laser current settings that we had a ~23dB gain from the 1550nm. We maxed out the powermeter at very low laser currents, and didn't continue to compare the emission from the 980 and the 1550 as the difference was substantial enough to clearly witness two photon absorption. Once roughly aligning the diode (OPA ~634nm) we could easily establish a signal from the 980nm laser where we were operating the laser at around .45 Amps that had a related power of over 300mw. We used the familiar piezo controls described in previous phases of the experiment to orient the diode in the 980 beam and maximized our signal at over 1mV, significantly higher than the peak potential we had seen using the 1550nm wavelength and the older (mystery) diode which had a maximum around 30uV.

Thus we took data and tried to refine ideas of how charge carriers operated within the junction and what we knew of two-photon absorption and band structure. We saw in our data that we had a much more defined PN junction that previous measurements suggest and we had definitely increased our resolution of the junction and the mapping of two-photon absorption within the junction.

Here is our first data:

Data1 a.gif

Data1 b.gif


We see this strange double peak where we may have expected the laser to provide us with a signal that would indicate a gaussian distribution of absorption across the junction. Next we turn down the intensity of the laser and repeat the same measurements with more attention to the behavior across the junction.

Here is our final data from the term:

SpringData2 3.gif SpringData2 1.gif SpringData2 2.gif

Interestingly we see this double peak while scanning the laser across the PN junction, namely the active region of the intrinsically doped region of the laser diode. We understand that in a PIN diode configuration, also described as a heterojunction - we see that the intrinsic region has the lowest bandgap energy which corresponds to the refractive index. In this way the active region of the laser diode acts as a waveguide, and this is the face of the laser that emits radiation. We consider our double peak somewhat as a result of the intensity of the laser. We know that as a non-linear optical phenomena, two photon absorption is very difficult to witness at low intensity but dominates at high intensity. We understand that perhaps when the beam is strictly on the active region, we are depleting the entire region of charge carriers and when we scan in nearby areas (such as overlapping a junction) there are more nearby available charge carriers to conduct. Thus this is the next phase of the experiment is to document the electric potential difference signal of the heterojunction by varying the intensity of the laser and documenting the behavior of this double peak. Our future measurements include continuing two-photon absorption, band edge measurements, exciton structure, etc...

DLTS Circuit Modifications and Overall Improvements

What we are seeking to measure in this experiment is the current output resulting from the loaded traps (lattice impurities) in the depletion region of a diode being emptied of charges as the loading pulse is shut off. Building on the work of the previous group, my goal was to see this current and report results based on varying both temperature as well as the semiconducting material being tested.

DLTS Circuit Modifications

The first major task was to ensure the DLTS circuit was functioning properly after recognizing the difficulties of the previous group. I isolated the DLTS circuit and tuned the variable resistor and capacitor such that there was no phase difference in the two legs of the original DLTS circuit pictured above. Once this was complete, I set about investigating the previous group's results and attempting to acquire a DLTS signal. Inital tests were done on the same 3 diodes used by the previous group, and results can be seen below for the P6KE30A diode:

suspected DLTS signal on P6KE30A diode

and under the same conditions for the first 3 diodes:

suspected DLTS results for 3 different diodes

Test program conditions: Function generator has a 30 MHz signal, +20 mV peak-to-peak amplitude, and a +1 V DC offset. The loading pulse (on the old pulse generator) has a +2.0V amplitude and 1ms length.


The suspected recombination current outputs from each respective diode appeared to be on the order of 0.02 - 2 mA. The above data was convincing enough to move onto the next major modification - attempting to amplify the output signal and investigate the nature of the traps in the depletion region of the P6KE30A diode.

Before that was done, a path length modification was installed. This amounted to mounting the diode on an elongated stretch of wire (approximately 240mm total increase) and balancing that path length on the other side of the circuit with an equal length of wire. This was done so that a very rudimentary temperature test could be done - dunking the entire diode in liquid nitrogen.

External Improvements

A new pulse generator, TTL switch, and 10db RF amplifier were tested and installed after the DLTS circuit to achieve two goals. Firstly, the new pulse generator would send two pulses: one to load charge traps in the diode by momentarily putting the diode into forward bias (as the previous setup was designed to do). Secondly, another pulse would be sent to the TTL switch and amplifier. This delayed pulse would be triggered to turn the amplifier on immediately after the diode-loading pulse was shut off. This would effectively amplify the output signal from the diode in the region of interest and allow the experimenter to see the output recombination current in greater detail. A schematic of the improvements can be seen below:

Modified DLTS Diagram

and a few images the updated setup:


A tunable phase shifter and 1 MHz high-pass filter will need to be installed in the future. At the end of my time on the project a working phase shifter could not be located, but a 1 MHz high pass filter has been created and can be found on the optics table near the experiment for later installation. The goal of the filter is simple noise reduction. The goal of the phase shifter is to ensure we are amplifying the output of the "decaying capacitor" that is the depletion region.

DLTS Data 2.0

After all the external improvements were implemented and determined to be working as designed, I resumed looking for the recombination current. Initial tests on the P6KE30A diode showed this output signal:

P6KE30A

Experimental conditions: 30 MHz RF signal, +20 mV peak-to-peak amplitude, +1 V DC offset. Loading pulse: +1.5 V for 20 us. NOTE: The pulse generator output amplitude should be approximately +3.70 V. The signal is split many ways, so this will yield approximately +1.5 V (which is setup to be read directly from the oscilloscope and is yellow in the above image).

It was at about this point that progress on the DLTS investigation ceased. Several anomalous signals were observed, including this example below, where green is the DLTS output, yellow is the loading pulse, and purple is the amplification pulse:

anomalous DLTS signal

The zero-crossing behavior and amplitude of the above signal couldn't be readily explained. It was determined that a resistor in the DLTS circuit had become disconnected and reconnecting this seemed to remedy the problem, but no further investigations could be carried out due to time constraints.

Future Directions

The first order of future business should be be to calculate an estimate of the observable recombination current in a silicon depletion region of similar volume to that of the P6KE30A diode. References on ideas for how to execute that calculation can be seen at the end of my final presentation, located here: File:Schachtner APL Final.pdf

Any future investigation should be predicated on a solid theoretical prediction for observable values. Most of my investment in the project was towards improving the initial circuitry setup and making sure it functioned properly. However, it is still unknown if the desired output was/is achievable given the current experimental setup.

The circuit could be improved right away and action could be immediately taken to install the high pass filter and phase shifter as mentioned previously.

Lastly, an intelligent and working implementation of the Cryostat designed by the previous group should be carried out so the temperature dependence of the recombination current can be accurately characterized.


Alex Cook and Nick Wogan Research Project

IV and CV Curve Measurements

Like other groups, we began by exploring the 'activation voltage' of a 365-1085-ND photodiode- that is, the voltage required in order for the p-n junction to conduct.

In order to do so, we applied varying bias voltages across the photodiode and calculated the current across a resistor in the following manner:

Setup1.png

The results came out quite good.

The following is a graph displaying the IV Curve of the Diode between 0-.8 V. Notice that we tested the IV Curve in three different lighting environments, in the dark with the lab lights off, in the ambient light of the lab, and underneath and intense flashlight.

IVcurves.PNG


The following is a graph displaying the IV Curve of the Diode between -40- 1 V. Notice the considerable difference between the flashlight and ambient light caused.

Fulliv.png

The measurements regarding the capacitance of a photodiode requires a slightly higher degree of sophistication. The following is a diagram of our methodology:

Setup2.png

Notice that we treat the photodiode in this case as part of a voltage divider. We treat is as an impedance producing device. Thus, the calculation to solve for the capacitance is as follows:

Calculationofcap.png

The following is a graph of our results:

Capacitance of Photo.png

The measurements did not turn out very well. Notice that between 0 and .3V, we found a negative capacitance, which is physically impossible. That said, the waveform of the results is quite good- capacitance increases quite quickly as a positive voltage is applied. We suspect there was an underlying problem in our methodology, but were unable to determine the root cause and, in order to complete our term goals, decided to move on rather than troubleshoot further.

Band Edge Measurement

We then began work on the principle focus of our research- the band edge measurement. A distinct type of semiconductor device is the laser diode. By applying a laser of sufficient wavelength to the p-n junction of the laser diode, we are able to produce a non-trivial voltage. This is, in essence, the photovoltaic effect. We sought to measure the wavelength (and therefore energy) necessary to stimulate this affect.

The following is a diagram of the experimental setup:

Setup3.png

The following is a graph of our preliminary results. Notice the high amount of noise, relative to the received signal. Also notice that the noise floated between each measurement.

Signalandnoise.png

The following is a graph of the signal with the noise subtracted. The measurement is okay. The band gap energy is seemingly delivered at a wavelength between 1554 and 1550nm.

Signal.PNG


Future Direction

We will be refining the band edge measurement in coming weeks and months. There are numerous improvements we can make to the experimental design, which should improve fidelity.

First, we will polish our fiber optic connections. We lose quite a lot of laser power in the 50/50 splitter and 95/5 splitter. We will also be applying a lens which should focus the laser beam on the active region of the PN junction. Increasing power delivered to the photodiode should make the band edge more apparent among the noise.

Secondly, we wish to reduce noise itself through the addition of lock-in amplification.

These additions should allow us to produce a far better measurement of the band-edge measurement.

With a good measurement achieved, we wish to study the excitonic conduction which occurs just below band-edge energy.