Difference between revisions of "Lock-in Amplifier"

From Advanced Projects Lab
Jump to: navigation, search
Line 18: Line 18:
 
There are multiple lock-in amplifiers in the lab, so the choice of which to use is yours.  There is one Model 5104 from Princeton Applied Research that may not be working properly, so you may want to avoid that device.  The Model 5202 from Princeton Applied Research was tested and seems to be working as it should.  Each lock-in amplifier has slightly different menus/controls, but the basics are all similar.
 
There are multiple lock-in amplifiers in the lab, so the choice of which to use is yours.  There is one Model 5104 from Princeton Applied Research that may not be working properly, so you may want to avoid that device.  The Model 5202 from Princeton Applied Research was tested and seems to be working as it should.  Each lock-in amplifier has slightly different menus/controls, but the basics are all similar.
  
Now that we have our equipment and a space in the lab to work we can set up our instruments.  Hopefully the function generator that you're using has two outputs, if it doesn't then another function generator might have to be used.  One of these outputs will be the reference signal and the other will be the lock-in's input signal.  Both signals should be sine waves for now and the frequencies of each can be set to whatever you desire, but if you're using the Model 5202 for example, then there is a corresponding setting on the lock-in for the reference signal frequency.  So you should make sure that dial is set to the right setting for whatever reference frequency you're using.  For now the reference signal and the input signal  
+
Now that we have our equipment and a space in the lab to work we can set up our instruments.  All devices must of course be powered but for now the lock-in doesn't have to be connected to the function generator.  Hopefully the function generator that you're using has two outputs, if it doesn't then another function generator might have to be used.  One of these outputs will be the reference signal and the other will be the lock-in's input signal.  Both signals should be sine waves for now and the frequencies of each can be set to whatever you desire, but if you're using the Model 5202 for example, then there is a corresponding setting on the lock-in for the reference signal frequency.  So you should make sure that dial is set to the right setting for whatever reference frequency you're using.  Something like 1 MHz should work.
 +
 
 +
The amplitudes of each signal are once again up to you, but the Model 5202 lock-in has a maximum input voltage of 5 Vrms, so make sure the amplitude of that signal is below that.  Something like 5 mVpp should work.  The reference signal on the other hand should be slightly larger.  Something like 1 Vpp, for example.
 +
 
 +
There are various settings on the front of the lock-in, some of which should be looked at.  The first dial on the left (Model 5202) represents a sensitivity, and for now can be set to something like 25 mV.  This can be changed later to make our equipment more sensitive.  There is also a setting for the reference phase (0 deg, 90 deg, 180 deg, etc.).  This is the phase difference applied to the reference signal when compared to the input signal.  For now it can be left to 0 degrees.  There may also be a setting for what waveform the reference will be.  We would want that set to a standard sinusoidal waveform.  The Model 5202 has some other Output settings as well, most of which can be left at the default.  The time constant can be left at the lowest value (ex. 10 ms).
  
  

Revision as of 16:24, 20 April 2016

Example.jpg

What is a Lock-In Amplifier?

A lock-in amplifier is a type of amplifier used to extract quiet signals out of noisy data. High quality lock-in amplifiers can extract signals up to a million times quieter than the surrounding noise. The output of a lock-in amplifier is a DC signal showing the strength of the signal to be extracted.

How Does a Lock-In Amplifier Work?

Conceptually, a lock-in amplifier works by exploiting the orthogonality of sinusoidal functions. Use of a lock-in amplifier requires a clear reference signal at the frequency of the signal to be extracted. This reference signal is multiplied by the noisy input signal and the product is integrated over a set time. When sinusoidal functions are multiplied together and integrated over a significant amount of time, the result will be zero unless the two sinusoidal functions have the same frequency. This is the before-mentioned orthogonality of sinusoidal functions. For the output of a lock-in, this means that the contributions of all signals not at the reference frequency will be attenuated very close to zero. The output is a DC signal showing the strength of the original input signal at the reference frequency.

From a circuits standpoint, a lock-in amplifier consists of a homodyne detector followed by an adjustable low pass filter. Traditional lock-in amplifiers used analog frequency mixers and RC filters for the demodulation, but modern devices are typically digital and use fast digital signal processing. The out of phase component of the signal that has the same frequency as the reference signal is also attenuated (sine functions are orthogonal to cosine functions of the same frequency), making lock-in amplifiers phase sensitive detectors. Sine and cosine demodulation is usually performed simultaneously (dual phase demodulation).


Try it in the Lab

Required Equipment: Lock-In Amplifier, Function Generator (two outputs), Oscilloscope, BNC Cables

There are multiple lock-in amplifiers in the lab, so the choice of which to use is yours. There is one Model 5104 from Princeton Applied Research that may not be working properly, so you may want to avoid that device. The Model 5202 from Princeton Applied Research was tested and seems to be working as it should. Each lock-in amplifier has slightly different menus/controls, but the basics are all similar.

Now that we have our equipment and a space in the lab to work we can set up our instruments. All devices must of course be powered but for now the lock-in doesn't have to be connected to the function generator. Hopefully the function generator that you're using has two outputs, if it doesn't then another function generator might have to be used. One of these outputs will be the reference signal and the other will be the lock-in's input signal. Both signals should be sine waves for now and the frequencies of each can be set to whatever you desire, but if you're using the Model 5202 for example, then there is a corresponding setting on the lock-in for the reference signal frequency. So you should make sure that dial is set to the right setting for whatever reference frequency you're using. Something like 1 MHz should work.

The amplitudes of each signal are once again up to you, but the Model 5202 lock-in has a maximum input voltage of 5 Vrms, so make sure the amplitude of that signal is below that. Something like 5 mVpp should work. The reference signal on the other hand should be slightly larger. Something like 1 Vpp, for example.

There are various settings on the front of the lock-in, some of which should be looked at. The first dial on the left (Model 5202) represents a sensitivity, and for now can be set to something like 25 mV. This can be changed later to make our equipment more sensitive. There is also a setting for the reference phase (0 deg, 90 deg, 180 deg, etc.). This is the phase difference applied to the reference signal when compared to the input signal. For now it can be left to 0 degrees. There may also be a setting for what waveform the reference will be. We would want that set to a standard sinusoidal waveform. The Model 5202 has some other Output settings as well, most of which can be left at the default. The time constant can be left at the lowest value (ex. 10 ms).


Useful Links

Informational PDF from Stanford Research Systems

Paper on Phase Sensitive Detection